What Are the Effects of External Carbohydrate Contaminants on Blood Glucose Measurements Made with a Glucometer?
Overview
Objectives/Goals The objective of this study is to determine if, and to what extent, carbohydrate contaminants on the skin raise blood glucose measurements made with a glucometer. Methods/Materials Carbohydrates (e.g., syrup and jelly), glucometer, syringes for measurement, and simulated blood (glucometer control solution). Experiment 1 simulates a typical life event: a finger is dipped in a carbohydrate and is quickly wiped off, a drop of simulated blood is applied, and the glucose level is measured with a glucometer. Experiment 2 is similar to Experiment 1, but the finger is wiped thoroughly. Experiment 3 examines tiny, quantified amounts of contaminants. Dilute carbohydrate solutions are created by dissolving small amounts of carbohydrates in water. Using a syringe, 0.01 mL samples are placed on a glass plate and allowed to dry. Using another syringe, 0.005 ml of simulated blood is applied to each sample and the glucose level is measured with a glucometer. Results In Experiment 1, the contaminants substantially raised the glucose measurement compared to the 125 mg/dL of uncontaminated simulated blood. For syrup the average was 298.3 mg/dL, for honey the average was 232.3 mg/dL, and for jelly the average was 279.3 mg/dL. Experiment 2 showed that even careful wiping was insufficient to prevent contamination. Here the average glucose level for syrup was 189.6 mg/dL. In Experiment 3, tiny, precisely-measured amounts of dried syrup solution produced an average measurement of 199.8 mg/dL, while dried glucose tablet solution produced an average of 455.4 mg/dL. These values quantify the amount of contaminant needed to cause measurement errors. Conclusions/Discussion Experiment 1 shows that small amounts of contaminant can even double the blood glucose measurement. Experiment 2 shows that even careful wiping is insufficient to eliminate the error in glucose measurements. Experiment 3 shows that a spherical speck of glucose tablet with a diameter of 0.77 mm increases the glucose by 260%, and a 0.8 mm diameter speck of syrup can increase it by 60%. The main lesson that can be gained from these experiments is that washing hands to remove contaminants is extremely important. Small amounts of contaminants can cause significant errors. The consequences of having a false glucose measurement can be disastrous for diabetics and physicians. An incorrect glucose measurement can result in the wrong treatment or diagnosis.
Summary statement
I showed that even tiny amounts of carbohydrates, such as honey and jelly, can cause substantial errors in blood glucose measurements made with a glucometer.
Help received
I developed the idea for the project and designed and conducted the experiments. I got help in measurement methods from Professor Stahovich in the Department of Mechanical Engineering at the University of California, Riverside. He also provided guidance on technical writing.
Competition history
- CSEF 2016
Resources
Related projects
CSEF · 2008
The Effect of Topical Contaminants on Blood Glucose Levels
CSEF · 2018
Out of Control: Blood Glucose Meter Accuracy
CSEF · 2019
How Much Glucose Is in That Drink?
CSEF · 2014
Surfing the Blood Glucose Wave: The Impact of Different Types of Sugar on Blood Glucose
CSEF · 2005
The Effects of Diet on Blood Glucose
CSEF · 2009
Sugar or Not: Cereal Effects on Blood Glucose
CSEF · 2014
Modeling a Blood Glucose Determination Using Surface Conductivity
CSEF · 2017
The Effects of Foods on the Blood Glucose of a Type 1 Diabetic and a Non-Type 1 Diabetic
Closest projects by meaning, across every fair and year in the corpus.
Browse more like this
Source: California Science & Engineering Fair public projects