Measuring Planck's Constant: Adventures in Watt Balance Land

CSEF · 2017 Physics & Astronomy Honorable_mention Award

Overview

Objectives/Goals Significantly reduce the measurement uncertainty for Plancks constant h attainable on a home-built watt balance by optimizing the laser detection system. Methods/Materials Inspired by the work of the mass metrology team at NIST, a fully functional watt balance instrument has been constructed. Custom coils were wound, a knife-edge main bearing was machined, electronics for detection and control were built. A floating battery powered supply was built to provide reverse bias on the photodiode detector, allowing higher intensity lasers to be evaluated. Four different lasers were evaluated for best performance in the shadow sensor detection system. The watt balance approach compares measurements made in force mode (static balance) and velocity mode. For a given system configuration, a measurement of h involved: calibrating the shadow sensor; n=10 replicate measurements of BLv in velocity mode and BLf in force mode. For force mode, 0-A-0 and A-B-A type measurements were compared. Results A baseline was established by measuring Plancks constant h with the system in stock configuration: h = 6.67 E-34 J s, precision 0.3%, accuracy 0.7%, uncertainty 0.7%. Uncertainty arising from the shadow sensor response was clearly the limiting factor in overall measurement uncertainty for h. Lasers for the shadow sensors were evaluated for sensitivity, linearity, beam profile, noise and drift. The best overall performance was achieved with a raw laser diode with no focusing optics. The calibration of the shadow sensor was compared with a simple numerical model for the diode laser Gaussian beam profile, with a 3rd order polynomial giving the best fit. With all improvements in place final results attained were: h = 6.64 E-34 J s, precision 0.2%, accuracy 0.2%, uncertainty 0.25%. Conclusions/Discussion A home-built watt balance apparatus to measure Plancks constant h has been constructed and refined. Multiple system enhancements (knife-edge bearings, alignment, shrouding) helped with reliable operation, and improvements to the shadow sensor (optimum laser, floating bias supply, cubic calibration fit) improved the overall measurement uncertainty for h by almost x4.

Summary statement

Measurement uncertainty for h was improved by almost x4 relative to NISTs published value for a home-built system, refining the shadow sensor response was the major factor in this improvement.

Help received

Thanks to my dad for subsidizing the necessary materials, helping me with the data acquisition and the poster graphics. Thanks to Gramps for gifting me the data logger and the voltage reference.

Awards (1)

  • Honorable Mention

Competition history

  • CSEF 2017 Physics & Astronomy · Entry S1817

Resources

Related projects

Closest projects by meaning, across every fair and year in the corpus.

Browse more like this

Source: California Science & Engineering Fair public projects

Save projects to your library

Sign in with Google to keep track of projects you find interesting, organized into folders. An account also raises your daily allowance for “Has this been done?”, and lets you create a key for the MCP server with a much higher limit than anonymous use. Browsing stays public.

Continue with Google